RFP QuestBeta
Awarded · ResultStage · contract

National Physical Laboratory

CAN: Optical Parametric Oscillator

Laboratory EquipmentCPV 38000000
Value£252k
Awarded16 Jan 2024
Published4 Apr 2024
RegionUK-wide
Outcome — awarded

This is a contract result notice, not an open opportunity. Details from the official award data.

Contract value in context
£252ktotal contract value
median £70k
this tender£0£588k

This is a large award for Laboratory & Precision Equipment — above three-quarters of comparable contracts. Based on 9,869 valued Laboratory & Precision Equipment tenders in our corpus.

The brief

The National Physical Laboratory requirement is for the supply of a laser source continuously tunable over the 350-1000 nm wavelength range.

The source is to be used to calibrate single-photon detectors (SPDs) in NPL's new traceable calibration facility.

The source is required to be pulsed at a variable repetition rate.

The pulse jitter must be short enough to allow timing jitter measurements of SPDs.

The pulse width must be short enough to enable measurement of the detection profiles of gated detectors.

The pulse-to-pulse energy variability must be low enough so that the intensity of individual pulses do not need to be measured, and the mean value represents a very good approximation to the energy of any individual pulse.

The technologies likely to fulfil these requirements are Optical Parametric Oscillators (OPOs) with a Second Harmonic Generator (SHG) module and a stable pump laser to allow a low pulse-to-pulse amplitude uncertainty, or an Optical Parametric Amplifier (OPA) with similar SHG and stable pump laser requirement.

The 350-1000 nm wavelength tuning range is required to be achieved without change of optics or re-alignment.

A pulse repetition rate between 1 MHz and 100 kHz is required (10 MHz-100 kHz is desirable).

The operational temperature range is at least from 15 C to 30 C.

Requirements

What the notice asks for

01

The National Physical Laboratory requirement is

The National Physical Laboratory requirement is for the supply of a laser source continuously tunable over the 350-1000 nm wavelength range.

02

The source is required to be pulsed

The source is required to be pulsed at a variable repetition rate.

03

The pulse jitter must be short enough

The pulse jitter must be short enough to allow timing jitter measurements of SPDs.

04

The pulse width must be short enough

The pulse width must be short enough to enable measurement of the detection profiles of gated detectors.

05

The pulse-to-pulse energy variability must be low

The pulse-to-pulse energy variability must be low enough so that the intensity of individual pulses do not need to be measured, and the mean value represents a very good approximation to the energy of any individual pulse.

Sentences from the notice that state an obligation, surfaced automatically and shown in the order they appear. Not an exhaustive list — always confirm against the tender documents.

Buyer intelligence

Make the case to bid

Reveal who to approach at National Physical Laboratory, and generate a go-to-market strategy from their news, accounts and people.

Sign up free to unlock buyer intelligence

Free to start. Named contacts come from published notices; the strategy is generated from the buyer’s news, accounts and people.

Source & provenance
OCID
b66c66a1-8acd-46f2-b32f-adf453957464
Stage
contract · Contract
Source
Contracts Finder
Buyer ref
tender_399537/1339113
View the original notice on Contracts Finder

Contains public sector information licensed under the Open Government Licence v3.0. Source data © Crown copyright.

Market context

Who wins this kind of work

The suppliers and buyers around this opportunity — drawn from official award data. Drag to orbit; click a node to explore.

Top suppliers & buyers in Laboratory & Precision Equipment

This view needs WebGL, which this browser has turned off.

National Physical Laboratory’s tender network