RFP QuestBeta
Awarded · ResultStage · award

University Of Edinburgh

Atmospheric gas analyser

Laboratory EquipmentCPV 38344000
ValueValue not published
Awarded
Published1 Apr 2026
RegionNationwide
Outcome — awarded
GeoGas Analytica

This is a contract result notice, not an open opportunity. Details from the official award data.

Who to contact
jpike2@ed.ac.uk
+44 1316502759

The procurement contact named on the official notice.

The brief

SUMMARY The University of Edinburgh Field Spectroscopy Facility requires a sensor system/analyser with the capability to measure N2O (nitrous oxide), CO2 (carbon dioxide) and CH4 (methane) concentrations at altitude/height simultaneously and in real time, at a detection limit in the <1 ppb for methane, <0.5 ppb for N2O and < 200 ppb for CO2.

The concentrations are required for flux measurements, and therefore the temporal frequency of measurements is important, measurement frequency needs to be no less than 1 Hz, ideally at 5 Hz.

PAYLOAD WEIGHT AND INTERFACE To enable use of the facility’s existing UAV equipment the sensor must be able to interface with a DJI M400 drone, and the total sensor weight must be under 7kg.

MEASUREMENT APPROACH Measurements of these gasses must be made simultaneously, to a synchronised timing and measurement frequency, and the measurement at a given timestamp should be taken from the same shared volume of air.

This should produce a single combined data series.

To give an example of the intention; At timestamp 1, a volume of air, V1, has been ‘ingested’, and will be tested by the analyser to give measurements of N2O(t=1), CO2 (t=1) and CH4(t=1).

At timestamp 2, (1 second in the future, if operating at 1 Hz), a new volume of air V2 is ingested and analysed to give measurements of N2O(t=2), CO2 (t=2) and CH4(t=2).

And so on, at the ‘Hz’ rating the device operates at.

Measurement must be conducted in-situ/in-flight at real time while the UAV operates; storing volumes of air for after-operation analysis would not be suitable.

CONCLUSION After a market evaluation it is believed there is only one solution which can fulfil the combined requirements – the Aeris Sensors MIRA Strato.

As such we intend to award via the Negotiated Procedure Without Prior Call for Competition, to their sole EU distributor, on the basis competition is absent for technical reasons.

Any suppliers who wish to query or challenge this notice should do so within 10 calendar days of its publication.

Value represents cost excluding VAT and import duties (if applicable).

Key requirements

What the supplier must deliver

01

The concentrations are required for flux measurements

The concentrations are required for flux measurements, and therefore the temporal frequency of measurements is important, measurement frequency needs to be no less than 1 Hz, ideally at 5 Hz.

02

To enable use of the facility’s existing

To enable use of the facility’s existing UAV equipment the sensor must be able to interface with a DJI M400 drone, and the total sensor weight must be under 7kg.

03

Measurements of these gasses must be made

Measurements of these gasses must be made simultaneously, to a synchronised timing and measurement frequency, and the measurement at a given timestamp should be taken from the same shared volume of air.

04

This should produce a single combined data

This should produce a single combined data series.

05

Measurement must be conducted in-situ/in-flight at real

Measurement must be conducted in-situ/in-flight at real time while the UAV operates; storing volumes of air for after-operation analysis would not be suitable.

Derived from the notice text — always confirm against the original documents.

Buyer intelligence

Make the case to bid

Reveal who to approach at University Of Edinburgh, and generate a go-to-market strategy from their news, accounts and people.

Source & provenance
OCID
ocds-h6vhtk-06478f
Stage
award · Awarded
Source
Find a Tender
Buyer ref
030291-2026
View the original notice on Find a Tender

Contains public sector information licensed under the Open Government Licence v3.0. Source data © Crown copyright.